8 PRECAUTIONARY INSTRUCTIONS
WARNING!
• This device should be used only under the continued supervision of a physician or
licensed practitioner.
• This device should be kept out of the reach of children.
• The patient and the device must be in sight of the therapist at all times.
• Should the use of this device lead to an undesirable event such as death or serious
injury to the user, the manufacturer and the competent authority of the Member
State MUST be informed without delay!
• In case of damage from transport is noticed, contact your local distributor.
DO NOT USE the device.
• Use of any other cable other than the supplied cable is strictly PROHIBITED as it
affects patient safety and the proper function of the device.
• Use of accessories, electrodes and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper
operation.
• Make certain that the unit is electrically grounded by connecting only to a grounded
electrical service receptacle conforming to the applicable national and local electrical
codes.
• Do not administer shortwave when a patient had an implanted lead in the past unless
you are sure that the implant and all leads in their entirety have been removed. Note
that leads are often left implanted after the implant is removed.
• It is considered safe to deliver a low dose (less than 5 Watts mean power) when there is
metal in the tissues. Metal plinths are generally considered acceptable when the applied
mean power is less than 5 Watts.
• Keep yourself informed of the contra-indications.
• In the interest of physiotherapist safety, it is recommended that once the machine has
been switched on, the physiotherapist and all other personnel should keep at least 1
meter from the operating machine, leads and electrodes. Pregnant physiotherapists or
others with concerns may want to ask a colleague to turn the CURAPULS 670 machine
on. The device will turn off automatically.
• Always perform a routine skin sensation test prior to a SWT application when a thermal
effect is anticipated (i.e. a mean power in excess of 5W).
• Do not operate the unit in an environment of short-wave or micro-wave diathermy use
as this can result in burns beneath the electrodes.
• There may be potential difficulties in ensuring electromagnetic compatibility in domestic
and those directly connected to a low voltage power supply networks due to conducted
as well as radiated disturbances.
• Always determine the dosage based on the patient’s perception of heat. Any sensation
greater than mild warmth could cause a burn.
• Before using the unit, verify the patient is not in contact with the unit, the electrode
connection cable, the electrodes, or other devices or metal objects.
• Do not use this device in so called “Wet Rooms” (hydrotherapy rooms).