Hamilton VisiFerm DO B Series Operating Instructions Manual

Typ
Operating Instructions Manual
VisiFerm
DO B
Sensors
Operating Instructions
Bedienungsanleitung
Important note
Copyright © 2013 Hamilton Bonaduz AG, Bonaduz Switzerland.
All rights reserved. The reproduction of any part of this document
in any form is forbidden without the express written agreement of
Hamilton Bonaduz AG.
Contents of this manual can be modified without previous
announcement. Technical modifications reserved. Greatest
possible care was used on the correctness of the information in
this manual. If errors should be discovered nevertheless, Hamilton
Bonaduz AG is pleased to be informed about it. Regardless of
this, Hamilton Bonaduz AG cannot assume liability for any errors
in this manual or for their consequences.
3
Operating Instructions VisiFerm
DO B Sensors
Table of Contents
TABLE OF CONTENTS
1. Introduction .............................................................. 4
2. Liability ...................................................................... 4
3. Intended use ............................................................. 4
4. Safety instructions ................................................... 5
5. Initial operation ......................................................... 5
6. Electrical connection ............................................... 6
Electrical connection of the 4–20 mA current interfaces ......7
Controlling 4–20 mA current interface signals
by pulse-width modulation ...................................................7
Examples of circuit arrangement .......................................... 9
Electrical connection for the ECS mode ............................... 9
Example of circuit arrangement .......................................... 11
Electrical connection for the digital RS485 interface .......... 11
Example of circuit arrangement .......................................... 12
7. Configuration and monitoring of the sensor .........13
8. Preparation for the measurement .......................... 13
9. Removal of the sensor ............................................14
10. Sterilization, autoclaving, CIP procedures ............ 14
11. Correction Mode for Cleaning In Place (CIP) .........15
12. Testing and maintenance .......................................16
Calibration ........................................................................... 16
Automatic standard calibration ........................................... 16
Product calibration .............................................................. 17
Calibration in ECS mode ..................................................... 17
Changing the ODO Cap ..................................................... 17
Self-diagnosis functions ...................................................... 18
13. Disposal ...................................................................18
14. Parts and Accessories ............................................ 19
15. Technical data ........................................................ 21
1. Introduction
This manual refers to the VisiFerm
DO B range of optical, sterilizable
oxygen sensors from Hamilton Bonaduz AG. These sensors are
compatible with all other components of the Hamilton Arc System,
including a complete family of intelligent sensors for pH, dissolved
oxygen and conductivity measurements in a process control.
*The VisiFerm
DO B 225 has, in reality, a shaft length of 215 mm. This is to
ensure optimal rinsing characteristics in replaceable armatures such as Retractex.
Hamilton VisiFerm
DO B sensors are quality products manufac-
tured according to the very latest research findings. Follow the
instructions given here to be sure of optimal safety and durability.
2. Liability
The liability of Hamilton Bonaduz AG is detailed in the document
“General Terms and Conditions of Sale and Delivery,” chapter 12.
Hamilton is expressly not liable for direct or indirect losses arising
from use of the sensors. It must in particular be insured in this con-
junction that malfunctions can occur on account of the inherently
limited useful life of sensors contingent upon their relevant applica-
tions. The user is responsible for the calibration, maintenance and
regular replacement of the sensors. In the case of critical sensor ap-
plications, Hamilton recommends using back-up measuring points
in order to avoid consequential damages. The user is responsible
for taking suitable precautions in the event of a sensor failure.
3. Intended use
VisiFerm
DO B sensors were developed for the measurement
of dissolved oxygen partial pressure, together with the following
derived measurement parameters:
•Percentbyvolumeoxygen
•Percentairsaturationbyoxygen
•Concentrationofoxygeninliquids
The main application for VisiFerm
DO B sensors is the in-line
measurement under demanding conditions in biotechnology,
brewery, beverage. The main characteristic that makes VisiFerm
sensors ideal for this application is their remarkable long-term
stability, even after frequent sterilizations and autoclavings.
Designation Order number P/N
VisiFerm
DO B 120 (ECS) 243090
VisiFerm
DO B 225* (ECS) 243091
VisiFerm
DO B 325 (ECS) 243092
VisiFerm
DO B 425 (ECS) 243093

IMPORTANT: These instructions must be read, under-
stood and followed by all staff using VisiFerm
DO B
sensors. Hamilton can assume no responsibility for
damage and operational disruption arising from failure to
observe these instructions.
5Operating Instructions VisiFerm
DO B Sensors
During development, special attention was paid to an optimum
sanitary design. All materials in contact with the solution meet
FDA requirements.
VisiFerm
DO B sensors provide electrical interfaces: ECS (clas-
sical electrochemical sensor), standard analog (4–20 mA) and
digital RS485 (Modbus RTU). These are built into each sensor
and are supported directly from the sensor head. The analog
4–20 mA and digital RS485 interfaces do not require any addi-
tional equipment such as amplifier or transmitter. Beside different
interfaces, the VisiFerm
DO B sensor also contains a tempera-
ture sensor (NTC 22 k). This temperature sensor only has to be
used for the compensation of temperature of the oxygen signal in
ECS mode (simulation of a classical amperometric oxygen sensor
by means of VisiFerm
DO B). The temperature sensor must not
be used for the regulation of the process temperature.
4. Safety instructions
VisiFerm
DO B sensors must be used for their intended appli-
cations, and in optimum safety and operational conditions. The
specifications (such as temperature or pressure) defined in the
section entitled ‘Technical data’ must not be exceeded under any
circumstances. Potential hazards can exist if the sensor is not
operated correctly or appropriately.
Assembly and maintenance must be performed only by trained
personnel.
Make sure that the PG 13.5 thread and the O-ring are not dam-
aged when screwing the sensor into the process. O-rings are
consumable parts which must be exchanged regularly (at least
once per year). Even when all required safety measures have
been complied with, potential risks still exist with respect to leaks
or mechanical damage to the armature. Wherever there are seals
or screws, gases or liquids can leak out undetected.
Before removing the sensor from its measurement setup, always
make sure that no process medium can be accidentally spilled.
The built-in temperature sensor can only be used for monitoring
the sensor conditions, not for controlling the process temperature.
5. Initial operation
This VisiFerm
DO B sensor has been carefully tested and is
ready for use. Always check a sensor for possible defects after
first unpacking it. In the unlikely event that you find a damaged
sensor, return immediately the VisiFerm
DO B in its original
packing to your Hamilton representative.
1 Socket head with VP 8 connector
2 Serial number
3 PG 13.5 thread
4 O-ring
5 Sensor shaft with stainless steel heat number
6 ODO Cap with oxygen-sensitive luminophore
(sensory element), ODO Cap serial number, part number
and stainless steel heat number.

1 2 3 4 5 6
The VisiFerm
DO B sensors are delivered directly from the factory
already pre-calibrated (according to the operational specifica-
tions). The integrated analog ECS and 4–20 mA interfaces and
RS485 digital interface (Modbus RTU) are configured according
to factory defaults. You can find full details, including serial num-
ber and most important specifications, on the certificate provided
with each sensor.
To be sure to avoid electrical damage to the sensor, carefully follow
all the instructions in the section entitled ‘Electrical Connection.
Before application of the sensor for measurement, monitoring or
regulation, be sure to first check its configuration by means of an
operational test.
6. Electrical connection
The VisiFerm
DO B sensor is fitted with
a VP 8 socket head. The eight golden
contacts are denoted as pin A… pin H. For
easy identification of each pin the head has
a mark between pin A and pin B.
For the easiest and safest connection of VisiFerm
DO B sensors,
always use Hamilton VP 8 cables. These are available in a range
of different lengths.
ATTENTION! In the ECS mode the sensor can
be demolished if you try to operate the sensor as a
4-20 mA sensor!
VP pin Function
A ECS mode: cathode (only in ECS mode)
B - 4 – 20 mA interface (mA interface #1).
- ECS mode: anode. Never connect in the ECS
mode with +2 V or more!
C Power supply: +24 VDC (7 to 30 VDC)
Start-up power: 1 W
Continuous power consumption: 600 mW
D Power supply: Ground
E Temperature sensor NTC 22 k for ECS mode
F Temperature sensor NTC 22 k for ECS mode
G RS485 (A)
H RS485 (B)
Casing Sensor shaft shielding. The outside shielding of
the VP cable should be connected to power
supply ground (Pin D).
A
B
C
D
E
F
G
H
ELECTRICAL CONNECTION
ATTENTION! To avoid humidity problems, make
sure that the ODO Cap is always attached fi rmly to
the sensor shaft, and that the O-ring between the
shaft and cap is undamaged.
7Operating Instructions VisiFerm
DO B Sensors
Electrical connection of the 4–20 mA
current interfaces
The 4–20 mA interface enables direct connection of the VisiFerm
DO B sensor to a data recorder, indicator, control unit or PCS
with analog I/O. Apart from the two wires required for the (tem-
perature compensated) 4–20 mA oxygen signal only a power
supply is required for the VisiFerm
DO B sensor (VP 8 pins C
and D, see below).
When using the 4–20 mA interface, pins have the following
designations with respect to VP cable conductor colors:
In an electromagnetically noisy environment, it is advisable to as-
sign the sensor’s shaft and/or VP cable shield to ground or earth.
This significantly improves noise immunity and signal quality.
The 4–20 mA interface is configured at the factory with the range,
value and unit for the measurements as indicated on the certifi-
cate. Follow the instructions in the section entitled ‘Configuration
and monitoring of the sensor‘ to adjust the sensor according to
the requirements of your application.
Controlling 4–20 mA current interface signals
by pulse-width modulation
Hamilton sensors use the method of pulse-width modulation
(PWM) to adjust the DC currents of the 4–20 mA interfaces cor-
responding to the measured values. In principle, the pulse width
(t
i
) of a rectangular signal with a constant frequency, the pulse
duty factor (t
i
/ T), is modulated and afterwards demodulated by
a low-pass filter to generate continuous analog DC signals. The
resulting value y
i
corresponds to the average of the PWM signal
(see Figures 1 and 2). The PWM-loads of the Sensors have
low-pass filters which are not able to eliminate all AC fractions of
VisiFerm
DO B VP pin VP 8 cable
4–20 mA three-wire interface, B Coaxial shield
functions as a current sink. black
It regulates the input current
according to the sensor
measurements. This interface
needs a separate power supply
(pin C). Factory default is assigned
to an oxygen measurement.
Power supply: +24 VDC C Coaxial core
(7 to 30 VDC) Red
Start-up power: 1 W transparent
Continuous power consumption:
600 mW
Power supply: Ground. D Coaxial shield
red
Sensor shaft (connected to Shield Cable shield
power supply ground). green-yellow
ELECTRICAL CONNECTION
ATTENTION! Ensure that the sensor is adjusted
at the 4-20 mA mode! If the sensor is confi gured in
the ECS mode, and if you use it as a 4-20 mA sensor,
then this operation can result in irreparable damages
in the sensor!
the used PWM frequency of 5 kHz due to technical impossibili-
ties. Therefore, the current signals of the 4–20 mA interfaces are
still overlaid by a certain AC current which should be masked
by lag smearing or input filters of the current input card of the
process control system (PCS). Recommended PCS settings
are a sampling rate below 3.5 kHz, an averaging over more than
1 s, and the use of galvanically separated inputs to avoid oscilla-
tions. It is also possible to use mathematical functions or isolating
amplifiers for signal processing filtering if necessary. For detailed
technical advice about suitable isolating amplifiers, please con-
tact Hamilton technical support.
Figure 1: Progress of a rectangular signal with a period T and a pulse
duration t
1
for the generation of an analog signal with the value y
1
.
Figure 2: Progress of a rectangular signal with a period T and a pulse
duration t
2
for the generation of an analog signal with the value y
2
.
ELECTRICAL CONNECTION
9
Operating Instructions VisiFerm
DO B Sensors
Analog interface 1 and 2
Galvanically not isolated, pulse width modulation with 5 kHz,
recommended PCS settings:
•Usegalvanicallyseparatedinputs
•Samplingrate<3kHzand n * 3.5 kHz
•Averageover>1s
Examples of circuit arrangement
Fig. A: Three-wire loop wiring diagram for the 4–20 mA interface.
Fig. B: The safest form of wiring, using an isolation amplifier. The figure represents
4–20 mA interface. (For detailed technical advice, please contact the technical
support at Hamilton.)
Electrical connection for the ECS mode
The ECS mode enables the simulation of an electrochemical
sensor. Thus a VisiFerm
DO B sensor can be connected to
classical measuring devices instead of amperometric oxygen
sensors (Clark cells). Furthermore only the power supply of the
VisiFerm
DO B sensor is necessary.
+
-
-
+
Sensor




4–20 mA
+24 VDC
0 VDC; Ground

+
-
-
+
input output
24 V
-
+
-
+
Sensor




4–20 mA
4–20 mA
+24 VDC
0 VDC; Ground

ATTENTION! The sensor generates the 4-20 mA



necessary for correct sensor functionality when used
in 4-20 mA setups. Figure 5 illustrates a solution for
the problem.
ELECTRICAL CONNECTION
When using the ECS interface, pins have the following designa-
tions with respect to VP cable conductor colors:
In an electromagnetically noisy environment, it is advisable to as-
sign the sensor’s shaft and/or VP cable shield to ground or earth.
This significantly improves noise immunity and signal quality.
The NTC temperature sensor attached to the pins E and F is
isolated from the remainder of the integrated electronics and is
used for the temperature compensation of the oxygen signal in
the measuring device. If the measuring device is not designed
for typical sterilizable oxygen sensors, then the measuring device
probably expects another type of temperature sensor. Therefore
the measuring device neither can compensate the temperature
depending oxygen signal nor display the temperature correctly.
Usually classical sensors are operated with a polarization voltage
between anode and cathode. This polarization voltage is supplied
by the measuring device. VisiFerm
DO B can be operated with
polarization voltages usual for electrochemical sensors. The sen-
sor is optimized for a polarization voltage of -675 mV.
For adjustment to different measuring devices and/or for simu-
lation of different amperometric sensors the current can be
adjusted between 0 and 500 nA.
VisiFerm
DO B VP pin VP 8 cable
Cathode A Coaxial core
black transparent
Anode B Coaxial shield
ATTENTION: In ECS mode black
never connect with a potential
higher as +2 V!
Power supply: +24 VDC C Coaxial core
(7 to 30 VDC), max. 1000 mW red transparent
Power supply: Ground D Coaxial shield
red
NTC 22 k E White
NTC 22 k F Green
sensor shaft (connect with the shield Cable shield
mass of the power supply) green-yellow
ATTENTION! Do not apply any high voltage (max.

of the sensor in ECS mode! Note: Only in 4-20 mA

order to operate the current interface!
ELECTRICAL CONNECTION
11
Operating Instructions VisiFerm
DO B Sensors
Example of circuit arrangement
Fig. C: Wiring diagram for the ECS interface.
Electrical connection for the
digital RS485 interface
The digital RS485 interface enables communication with Visi-
Ferm
DO B sensor for performing measurements, for calibrating
the sensor and for changing the sensor’s configuration param-
eters. VisiFerm
DO B sensors are always connected to digital
controlling devices as a Modbus slave. To function, they require
a power supply (VP 8 pins C and D, see below). The section
entitled ‚Configuration and monitoring of the sensor‘ describes
operation in digital mode.
By using the correct access password, the Arc System operator
can adapt Arc sensors to many tasks by:
• Switchingbetweenthe4–20mAandECSinterfacemode.
• Scaling(configuring)the4–20mA,orECSinterface.
• Selectingthemeasuredparameter:
- DO: % oxygen air saturation (%-sat), volume % of oxygen
(%-vol), mg/l or ppm, ug/l or ppb.
- Temperature T: °C.
• Compensatingfortheprocesspressure,forcalculationof
volume percentage values (%-sat, %-vol).
• Compensatingforthesalinityofthemeasuredsolution,
for calculation of the concentration values (mg/l or ppm,
µg/l or ppb).
In addition, operators can read sensor information from the
RS485 interface such as:
• Thesensor’sserialnumber,partnumber(P/N)and
manufacturing number (WO).
• Thesensor’sfirmwareversion.
• Thesensor’sstatus(e.g.operationhours,numberof
cleanings and sterilizations, warnings and errors).
Additional information:
The Modbus RTU communication protocol corresponds to the
Modbus-IDA standard (see www.modbus.org). VisiFerm
DO B
uses an open register set developed by Hamilton. Additional
information about the register content and structure can be found
at www.hamiltoncompany.com.
Sensor
 Cathode



 Anode

Temperature

+
-
+24 VDC
0 VDC; Ground
Transmitter
ATTENTION! Because all sensors are delivered
with factory-default settings, each sensor must be
configured for its specific application before first
use (see the section entitled ‘Configuration of the

ELECTRICAL CONNECTION
The pins for digital the RS485 interface have the following
designation with respect to VP cable conductor colors:
In an electromagnetically noisy environment, it is advisable to
connect the VP cable shield to the ground. This significantly
improves noise immunity and signal quality.
Example of circuit arrangement
Fig. C: Wiring diagram for the RS485 interface.
Fig. D: Multi-drop bus wiring for the Modbus two-wire mode.
Each sensor functions as a Modbus slave.
The serial Modbus connection between the RS485 port of the
master and the corresponding interfaces of the sensors has to
be ensured according to the EIA/TIA RS485 standard. Only one
sensor can communicate with the master at any time.
NOTE: In the connection scheme shown above, each
sensor must have the unique Modbus device address for
proper communication.
VisiFerm
DO B VP pin VP 8 cable
Power supply: +24 VDC C Coaxial core
(7 to 30 VDC), max. power red transparent
consumption 1 W.
Power supply: Ground. D Coaxial shield red
RS485 (A) G Yellow
RS485 (B) H Brown
Sensor shaft (connected to the Shield Cable shield
ground of the power supply) green-yellow.
+
-
Sensor




+24 VDC
0 VDC; Ground


A
B
Ground
RS 485


13Operating Instructions VisiFerm
DO B Sensors
7. Configuration and
monitoring of the sensor
Two options are available for the digital configuration and
monitoring of a sensor:
1. Personal computer or notebook. The following additional
equipment and software is also required:
• HamiltonUSB-RS485Modbusconverter(P/N:242411)
• Hamiltonfreeware‘HDM(HamiltonDeviceManager)‘
available at www.hamiltoncompany.com (follow the
instructions in the Configurator’s user guide for the
installation and operation).
• DemoCable(P/N:355194-xx).Thiscableincludesapower
adapter to supply the sensor with operation power, and a
plug to connect the two RS485 conducting wires (yellow
and brown) to USB-RS485. If you use a standard VP 8
cable, you must supply the sensor with an external power
source (pin C: 24 VDC; pin D: Ground).
2. Hamilton Arc View Handheld Package (P/N: 242180). The
handheld package represents an ideal desktop solution for
the Arc sensor management. The Arc View Handheld included
in the package is a compact mobile wireless device with long
battery life and broad functionality. When using the Handheld
as a mobile wireless device, each Arc sensor requires an Arc
Wi Sensor Adapter (P/N: 242170).
The digital RS485 interface of the VisiFerm
DO B sensor can be
accessed by operators working within a three-tier hierarchy. Op-
erator hierarchy levels and factory default passwords are shown
in the table below.
The Users can read basic data from sensor. Administrators can
also calibrate sensors. Specialists can calibrate and configure
sensors, and can see all data.
8. Preparation for the
measurement
In contrast to electrochemical oxygen sensors, VisiFerm
DO B
sensors work independently of installation position orientation.
Prepare the sensor for measurements as follows:
1. Remove the protective caps from the VP head and from the
ODO Cap.
2. Make sure that the sensor is configured as required. If in
doubt, test as described in the section entitled ‘Electrical
connection: RS485 Modbus interface.
3. Check the condition of the sensor and ODO Cap.


Operator status Operator Level Password
User U Not required
Administrator A 18111978
Specialist S 16021966
4. Make sure that the ODO Cap is attached firmly to the
sensor shaft, and that the O-ring between the shaft and
cap is undamaged.
5. Installation of the sensor (thread: PG 13.5).
6. Attach the VisiFerm
DO B sensor according to the section
‘Electrical Connection‘ in the desired configuration (analog
4–20 mA interface, digital RS485 interface or both).
The sensor signal stabilizes itself within few minutes. The sensor
is factory pre-calibrated and ready for use. However, if an on-site
calibration is necessary, a calibration can be performed via
RS485 by means of the Arc View Handheld, personal computer
or VisiCal
(P/N: 242410).
9. Removal of the sensor
Before removing any sensor from a process, make sure that no
process medium can escape during removal.
10. Sterilization, autoclaving,
CIP procedures
VisiFerm
DO B sensors are carefully designed to withstand the
cleaning procedures normally used in biotechnological industries
without the need for special precautions. Nevertheless, as is the
case with all sensors, it is seen that frequent cleanings or steriliza-
tions with steam or hot and aggressive solutions lead to short-
ened life span of the sensor. Although for a number of reasons
(variations in composition of cleaning agent, temperature gradient
and so on) it is not possible to give general life span figures. With
steam sterilizations and autoclaving for 30 min at a temperature
of 130 °C, a life span of more than 50 cycles per exchangeable
ODO Cap is common.
The contacts must be clean and dry before sensor is connected
to the cable.


NOTE: Due to the integrated optoelectronics the sensor
has an intrinsic warming which stabilizes itself within a few
minutes. If the sensor is not dipped into the liquid, then it
displays a reading that is slightly too high.
-
tion and pull out the sensor. Never turn the sensor at
the socket head because by doing so you can loosen
the ODO Cap and fluid can reach the interior of
the sensor.
ATTENTION! The sensor can communicate over

-

the luminophore, no oxygen measurement is performed

the sensor cannot be used for control or monitoring.
15Operating Instructions VisiFerm
DO B Sensors
11. Correction Mode for
Cleaning In Place (CIP)
The CIP correction mode is designed to compensate for a mea-
surement shift which may be created during the process. It has
been designed for environments, such as breweries, which may
not allow calibration of a sensor after every CIP cycle.
Two optional and independent CIP correction modes have been
implemented in the VisiFerm
DO B:
•CIPcompensationmode
 •Brewerymode
The CIP compensation mode enables the definition of a fix
offset (phase shift in degrees) which corrects a CIP-driven sensor
shift. It may be used where the CIP process is reproducible.
The offset is entered as the average measurement shift observed
after due to a given CIP process, and used to correct the mea-
surement value after each CIP cycle. Please refer to the conver-
sion table to translate the ppb offset in phase degree for 100%
ODO Cap quality.
The brewery mode automatically corrects small sensor drifts
after CIP and during measurement. It is active only if the DO
reading is lower than 150 ppb, and the temperature is lower than
20 °C. The VisiFerm
DO B detects rapid DO intrusion higher
than 150 ppb.
Note: An accurate CIP compensation requires
reproducible CIP process.


Offset in (ppb) Phase shift (°)
5 0.05
10 0.1
20 0.2
30 0.3
50 0.5
80 0.8
100 1.0
200 1.9
500 4.6
ATTENTION! Slow intrusion of oxygen traces

ATTENTION! The sensor must be calibrated after

defi ned in the technical specifi cations.
12. Testing and maintenance
Calibration
The VisiFerm
DO B sensors provide two kinds of sensor cali-
bration: automatic standard calibration, and product calibration.
The automatic standard calibration is performed over RS485
interface using the Arc View Handheld, VisiCal
or the HDM
(Hamilton Device Manager) freeware (see the sections entitled
Configuration and monitoring of the sensor‘ and ‘Accessories‘).
For the product calibration use the Arc View Handheld, or the
HDM (Hamilton Device Manager) freeware (VisiCal
is not sup-
porting the product calibration.)
Automatic standard calibration
The VisiFerm
DO B sensor is calibrated at two points: in air and
in an oxygen-free environment. During calibration, the sensor
examines the correctness and stability of the oxygen and tem-
perature signals. However, the recalibration of the VisiFerm
DO B
is very rarely necessary for most applications.
Calibration at point 1:
1. Set an appropriate operator level (Administrator or Specialist).
2. Immerse the sensor into an oxygen-free environment.
3. Let the system equilibrate. Guarantee stable conditions for at
least three minutes. For greater measurement accuracy insure
that temperature difference between calibration medium and
process medium is minimal.
4. Execute the calibration at point 1.
5. If all conditions are met, the sensor confirms the calibration im-
mediately. The calibration curve of the sensor is now defined by
the recently calibrated point 1 and the prior calibration at point 2.
Calibration at point 2:
1. Set an appropriate operator level (Administrator or Specialist).
2. Leave the sensor for at least three minutes under stable con-
ditions in ambient air or in oxygen saturated medium.
3. Execute the calibration at point 2.
4. If all conditions are met, the sensor confirms the calibration
immediately. The calibration curve of the sensor is now defined
by the recent calibration at both points: 1 and 2.
If you experience measurement value shifts as small as a few
percent within a period of a few days, suspect damage to the
luminophore caused by the measurement or cleaning medium.
Should this occur, consider whether more frequent exchange of
the ODO Cap is acceptable, or whether a different approach to
measurement (for example, using a classical OxyFerm sensor)
should be taken.
NOTE: The sensor requires a preheating period of 10 to 15 min-
utes after it is switched on. Although measurements are possible
during this time, you must wait until preheating is complete to be
able to calibrate the sensor optimally.
NOTE: The concept behind VisiFerm
DO B sensors enables
calibration and configuration in the lab before use in the process
control. Another calibration for the installation in the process
setup is not required.

17
Operating Instructions VisiFerm
DO B Sensors
Product calibration
The product calibration is an in-process calibration procedure in
order to adjust the measurement to specific process conditions, or
in case the sensor cannot be removed for the standard calibration.
Product calibration is an additional calibration procedure to a
standard calibration. Product calibration adapts the standard
calibration curve to the process conditions in force at the time
of product calibration. If product calibration is activated, the
VisiFerm
s calibration curve is calculated from the data of last
calibration at point 1 and from the data of the product calibra-
tion. In order to restore the original standard calibration curve, the
product calibration can be cancelled at any time. A new standard
calibration cancels a product calibration as well.
Actions:
1. Set an appropriate operator level (Administrator or Specialist).
2. Perform an initial measurement while taking a sample from the
process. The data of the initial measurement are stored in the
sensor.
3. Perform a laboratory measurement of the sample at the same
temperature as it was measured in the process.
4. Assign the laboratory value to the value of the Initial measure-
ment. This new DO value is accepted and instantaneous
active, if the difference between initial measurement and
laboratory values is not greater than 20%-sat units.
Calibration in ECS mode
The calibration in ECS mode takes place like in classical sensors
in the measuring device at which the sensor is connected. For
this, consult the operation instructions of the measuring device!
NOTE: The concept of VisiFerm
DO B sensors enables the
pre-calibration using the Arc View Handheld, Visical
or the HDM
(Hamilton Device Manager) freeware and then a direct use at
classical measuring devices without a new calibration at the
measuring device. This is possible, if the sensor was recently
calibrated and in the configuration of the sensors ECS interface an
air current of 60 nA was adjusted. The measuring device has set
the air current at 25 °C on 60 nA correspondingly.
Changing the ODO Cap
The exchange of ODO Cap is performed very easily:
Unscrew the old ODO Cap from the shaft.

O-ring
unscrew
NOTE: The product calibration is possible for DO values
in the range of 2 to 55 %-vol (20 – 550 mbar pO
2
).
NOTE: If the ODO Cap is mounted very firmly on the shaft, and
if you cannot obtain a good grip on the stainless steel with your
fingers, a silicone tube between your fingers and metal may supply
a better grip.
Examine the small O-ring that seals the ODO Cap to the sen-
sor shaft. Exchange the O-ring, if any traces of wear are seen. A
replacement O-ring is included with each replacement ODO Cap.
Screw the new ODO Cap onto the sensor shaft again. Make
sure that the gap between the shaft and cap is closed, and that
the sealing effect of the O-ring under it is therefore guaranteed.
In order to enable traceability (as required by the Federal Drugs
Administration) note the serial number of the new ODO Cap.
Examine the measurement values of the sensor in air, and if
necessary, in an oxygen-free medium. If the measurement values
deviate significantly from operated value, perform a calibration.
Self-diagnosis functions
VisiFerm
DO B sensors provide a self-diagnosis functionality to
detect and identify the most common sensor malfunctions. Both
interfaces can be used for warning and error messaging. The
analog 4–20 mA interface can be configured according to the
NAMUR recommendations to indicate an abnormal event. The
RS485 interface provides a variety of the indications based on the
fault code.
Use the Arc View Handheld and Arc Wi Sensor Adapter for moni-
toring of the sensor status and troubleshooting. These devices
detect and highlight the sensor condition according to the fault
code. The Arc View Handheld displays the text messages that are
appropriate for the type of malfunction.
The following types of messages are provided by the self-diagnosis
function:
•Warning(alarm):
- DO readings upper/lower range, unstable.
- Temperature readings upper/lower, unstable.
- Calibration recommended.
- Calibration upper/lower, out of range, or unstable.
- Replace ODO Cap.
- 4–20 mA output out of range, unstable
- Supply voltage upper/lower range.
•Error(failure):
- ODO Cap missing (reading failure).
- Temperature out of range.
- Temperature sensor defect (reading failure).
- Supply voltage out of range.
13. Disposal
The design of Hamilton sensors optimally consid-
ers environmental compatibility. In accordance with
the EC guideline 2002/96/EG Hamilton sensors that
are worn out or no longer required must be sent to
a dedicated collection point for electrical and electronic devices,
alternatively, must be sent to Hamilton for disposal. Sensors must
not be sent to an unsorted waste disposal point.

19Operating Instructions VisiFerm
DO B Sensors
14. Parts and Accessories
Arc View Handheld with Dock
Arc Wi Sensor Adapter for wireless communication
VisiCal

Order number Describtion
242427 ODO Cap P0
243510 ODO Cap P1
242411 USB-RS485 Modbus Converter
243498 Arc Wireless Converter ModBus
242180 Arc View Handheld Package
355194-xx Demo Cable
(1m, open end, with plug power pack)
-01 (EU), -02 (CH), -03 (US),
-04 (UK), -05 (AUS)
242170 Arc Wi Sensor Adapter
242410 VisiCal
355217 Sensor Cable VP 8, 1 m
355218 Sensor Cable VP 8, 3 m
355219 Sensor Cable VP 8, 5 m
355220 Sensor Cable VP 8, 10 m
355221 Sensor Cable VP 8, 15 m
355222 Sensor Cable VP 8, 20 m
USB-RS485 Modbus Converter
Sensor Cable VP 8

  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

Hamilton VisiFerm DO B Series Operating Instructions Manual

Typ
Operating Instructions Manual

in anderen Sprachen