Invacare IOH200AW Benutzerhandbuch

Typ
Benutzerhandbuch
Invacare®HomeFill®II
ModelIOH200AW
enCompressor
UserManual................................3
csKompresor
Návodkobsluze............................29
frCompresseur
Manueld'utilisation..........................57
deKompressor
Gebrauchsanweisung.........................87
itCompressore
Manualed’uso...............................117
plCompresor
Instrukcjaobsługi............................147
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2014Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
WD–40and3–IN-ONEareregisteredtrademarksofWD-40Company.
UpdatedversionsofthismanualareavailableonyourregionalInvacarewebsite.Foralistofregional
Invacarewebsites,refertotheaddresseslistedattheendofthismanual.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
2Safety..........................................6
2.1GeneralGuidelines.............................6
2.2LabelLocation.................................9
2.3DealerInformation.............................11
2.4Disposal.....................................11
2.5EMCInformation...............................12
3Set-Up.........................................13
3.1Unpacking....................................13
3.2CompressorInitialSetupChecklist..................13
3.3ChoosingaLocation............................14
3.4TransportingtheCompressor.....................14
3.5InstallingthePowerCord.........................15
3.6ConnectingtheCompressortotheConcentrator.......15
4Usage..........................................16
4.1CompressorOperationChecklist...................16
4.2CylinderPrefillInspection.........................16
4.3TheHydrostaticTestingDate......................17
4.4ExternalExamination............................17
4.5Connecting/DisconnectingtheCylinderto/fromthe
Compressor.................................17
4.5.1ConnectingtheCylindertotheCompressor........17
4.6TurningtheCompressorOnandOff................19
5Maintenance....................................21
5.1MaintenanceInfo...............................21
5.2CleaningtheCompressorFilter....................21
5.3CleaningtheCabinet............................21
5.4ReplacingtheFuse..............................22
6Troubleshooting.................................23
6.1IndicatorLights................................23
6.2Troubleshooting...............................24
7TechnicalData..................................27
7.1TechnicalDescription...........................27
7.2CylinderFillTimes..............................27
7.3TypicalProductParameters.......................28
Invacare®HomeFill®II
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
IMPORTANT
Indicatesahazardoussituationthatcouldresultin
damagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
ReadtheManual
AlternatingCurrent
TypeBFequipmentAppliedParts
Powerswitch“ON”
Powerswitch“OFF”
DONOTsmoke
ClassII,DoubleInsulation
Keepdryintransportorstorageandduringuse
Fuse
Oxidizer
4
1145804-E
General
Non-flammablegas,class2type(oxygen)
EuropeanRepresentative
Manufacturer
TransportandStorageTemperature
TransportandStorageHumidity
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheCE
declarationofconformity.
1.2IntendedUse
TheintendeduseofanInvacareHomeFillcompressorisanaccessory
devicetoanInvacareoxygenconcentratorthatenablesthepatient
torefillcustomportableoxygencylindersforpersonaluse.
Thisproductisintendedforsinglepatientuseintheindoorhomecare
environmentonlyandisnottobeusedforbulkcylinderrefilling.
Nospecificproductknowledgeortrainingisrequired,otherthan
whatiscontainedinthismanual,tooperatetheproduct.
Itisnotintendedforsupplyingoxygentherapyduringrefilling.
Thisproductisintendedtobeusedbyorunderadultsupervision
onlyafterreadingandunderstandingtheinstructionsandwarnings
ofthisusermanual.
NOTICE
TheHomeFillIIcompressorisforusewithspecially
designedInvacareoxygenconcentratorsthatare
HomeFillcompatible.
TodetermineifyouroxygenconcentratorisHomeFillcompatible,
examinethelowerrightcorneroftherearoftheoxygen
concentratorAforanoutletfittingB.
PlatinumTopView
Perfecto
2
RearView
1145804-E5
Invacare®HomeFill®II
2Safety
2.1GeneralGuidelines
WARNING!
Thissectioncontainsimportantinformationforthesafe
operationanduseofthisproduct.
Readandunderstandtheseinstructions.DONOTuse
thisproductwithoutreadingandunderstandingthese
instructions.Ifyouneedhelpwiththeseinstructions,
contactahealthcareprofessionaloranInvacare
dealer.Otherwiseinjuryordamagemayoccur.
ACCESSORIESWARNING
Invacareproductscanonlybeusedwithaccessories
madebyInvacare.AccessoriesnotmadebyInvacare
havenotbeentestedbyInvacare.Accessoriesnot
madebyInvacareshouldnotbeused.
DANGER!
RiskofFire
UsersMUSTNOTSMOKEwhileusingthisdevice.
Keepallmatches,lightedcigarettesorothersources
ofignitionoutoftheroominwhichthisproductis
located.Textilesandothermaterialsthatnormally
wouldnotburnareeasilyignitedandburnwithgreat
intensityinoxygenenrichedair.Failuretoobserve
thiswarningcanresultinseverefire,propertydamage
andcausephysicalinjuryorDEATH.
Checkallpartsforshippingdamage.Incaseofdamage,DO
NOTuse.ContactInvacareforfurtherinstruction.
Inordertoensurethesafeinstallation,assemblyandoperationofthe
HomeFillIICompressortheseinstructionsMUSTbefollowed.
CAUTION!
Statutorylawcanrestrictthisdevicetosaleorrental
byoronorderofaphysician,oranyotherpractitioner
licensedbythelawofthegovernmentagencyinwhich
he/shepracticestouseorordertheuseofthisdevice.
WARNING!
Theuseofthisdeviceislimitedtorefillingaspecific
patient’sownoxygencylinders.Thedesignated
cylindersMUSTbeusedONLYbythespecificoxygen
patientandarenottobedistributedtoanyother
individualforanypurpose.
Nootheruseofthisdeviceisindicated.
Themaximumoxygenconcentratorflowrateforpatientuse
islistedinthetablebelow.Otherwise,theoxygenlevelto
thecompressorwillbe<90%andthecompressorwillnot
fillthecylinder.
InvacareOxygen
Concentrator
MaximumOutputFlowfor
PatientUse
Platinum5
2.5L/min
Platinum95L/min
Perfecto
2
3L/min
Patientswithahigherprescribedflowshouldnotrefilla
cylinderwhilereceivingsupplementaloxygentreatment
fromtheoxygenconcentrator.
61145804-E
Safety
WARNING!
NEVERblocktheairopeningsoftheproductorplace
itonasoftsurface,suchasabedorcouch,wherethe
airopeningmaybeblocked.Keeptheopeningsfree
fromlint,hairandthelike.
DONOTuseoutdoors.Indooruseonly.
NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor
atmosphere.
DONOTdisassemble.Nomodificationofthis
equipmentisallowed.Theunitisprovidedwith
doubleinsulationandhasnointernaluserserviceable
parts.Ifanyserviceisrequired,contactInvacare
TechnicalService.
Ifthecompressorhasadamagedcordorplug,ifit
isnotworkingproperly,ifithasbeendroppedor
damaged,orsubmersedinwater,callaqualified
technicianforexaminationifrepairisneeded.
IfanyleakageoftheoxygencylinderisdetectedDO
NOTattempttousethecylinder.Turntheflow
selectorknobontheregulatorto“Off”.Ifleakage
persists,placethecylinderoutdoorsandnotifyyour
HomeHealthCareProviderorServiceRepresentative
ofthiscondition.
ChildrenshouldALWAYSbesupervisedaroundthe
HomeFillIIunit.WhenusedwiththeHomeFilltable
orsimilarelevatedsupportsurfaces,achildcouldpull
overanysuchtablecausingtheunittofall.Failure
toobservethiswarningmayresultinpersonalinjury
orproductdamage.
WARNING!
DONOTallowoilfromyourhandsorothersources
tocomeintocontactwiththecompressorcoupling
orcylinderfittings.Thesesubstancesmaybecome
flammableinthepresenceofoxygenandcauseinjury.
Avoidtouchingthecylinderfittingwheneverpossible.
Exampleofitemstoavoidarehandcreamsorlotions,
cookingoils,suntanoilsorsunblocksandsimilaroily
products.IfyouusethesetypesofproductsInvacare
suggestsyouwashyourhandspriortousingthe
HomeFillunit.Ifthecylinderfittingshouldcomein
contactwithsuchsubstancesitshouldbecleaned
offwithadampclothpriortoconnectingittothe
compressor.
NEVERoilorlubricatethecompressorcoupleror
cylinderfittings.NEVERuseanypenetratingoil
orlubricantsuchasWD-40®or3-in-1Oil®.If
connectiondifficultyisexperiencedverifythatthe
cylinderfittingisbeinginsertedstraightdowninto
thecompressorcoupler.Ifcontinueddifficultyis
experiencedcontactyoursupplierforassistance.
Keeptheoxygentubing,cord,andunitawayfrom
heatedorhotsurfaces,includingspaceheaters,
blankets,stovesandsimilarelectricalappliances.
1145804-E7
Invacare®HomeFill®II
WARNING!
TheHomeFillIIisequippedwithahighpressurerelief
valvetoensuretheuser’ssafety.Whenactivated,this
safetyfeaturewillmakeanextremelyloudnoise.Ifthis
noiseoccurs,turntheunitOffandcontactyourHome
HealthCareProviderorServiceRepresentative.
Beforemovingorrepositioningeitherthecompressor
orconcentrator,ALWAYSpushthepowerswitchto
theOff(O)position,disconnecttheACpowercords
andtheinterconnecthosebetweenthecompressor
andconcentrator.Failuretodosomayresultin
damagetotheunitorpersonalinjury.
81145804-E
Safety
2.2LabelLocation
1145804-E9
Invacare®HomeFill®II
A
SerialNumberLabel
B
SpecificationLabel
101145804-E
Safety
2.3DealerInformation
Invacarerecommendsleavingafullcylinderofoxygenwith
thepatientaftersettinguptheHomeFillIIcompressor.
BeforeleavingaHomeFillIICompressorwithauser,thefollowing
checklistMUSTbecompleted:
qMakesurethepressuretestoncylinderhasnotexpired.
qChecktheEMCsectionforoperationalinformationandpossible
RFissues.Referto2.5EMCInformation,page12
.
qInstructtheuseronthesafeoperationoftheconcentratorand
reviewallwarnings.
qInstructtheuseronthesafeoperationoftheHomeFillII
compressorandreviewallwarnings.
qLeaveafullcylinderofoxygenwiththeuser.
qLeaveacopyofallwarningsandfillinginstructionswiththeuser.
2.4Disposal
Recycle
Thisproducthasbeensuppliedfromanenvironmentallyaware
manufacturerthatcomplieswiththeWasteElectricalandElectronic
Equipment(WEEE)Directive2002/96/CE.
Thisproductmaycontainsubstancesthatcouldbeharmfultothe
environmentifdisposedofinplaces(landfills)thatarenotappropriate
accordingtolegislation.
The'crossedoutwheeliebin'symbolisplacedonthisproductto
encourageyoutorecyclewhereverpossible.
Followlocalgoverningordinancesandrecyclingplansregarding
disposalofthedeviceorcomponentsnormallyusedinoperation.
Thedevicedoesnotgeneratewasteorresidueinoperation.DO
NOTdisposeofthedeviceinthenormalwastestream.Any
accessoriesnotpartofthedeviceMUSTbehandledinaccordance
withtheindividualproductmarkingfordisposal.
Pleasebeenvironmentallyresponsibleandrecyclethisproductatits
endoflife.ContactInvacareoryourserviceproviderforproduct
recyclinginformation.RefertotheregionalInvacare.
1145804-E
11
Invacare®HomeFill®II
2.5EMCInformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccording
totheEMCinformationinthismanual.
ThisequipmenthasbeentestedandfoundtocomplywithEMC
limitsspecifiedbyIEC/EN60601-1-2forTypeBFequipment.
Theselimitsaredeterminedtoprovidereasonableprotection
againstelectromagneticinterferenceinatypicalhomehealthcare
environment.Thisequipmentisnotintendedforuseinaprofessional
healthcareenvironment.
PortableandmobileRFcommunicationsequipmentcanaffectthe
operationofthisequipment.Thecompressormaybestackedontop
onanInvacareHomeFillcompatibleoxygenconcentratorduring
setup.Thecompressoroperationshouldbeobservedtoverify
normalcylinderfillingineithersituation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels
ofelectromagneticemissionspermittedbytheabovestandard.To
determineiftheemissionsfromthecompressorarecausingthe
interference,turnthecompressorOff.Iftheinterferencewiththe
otherdeviceoperationstops,thenthecompressoriscausingthe
interference.Insuchrarecases,interferencemaybereducedor
correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
devices.
Connecteitherlinepowereddevicetoadifferentelectrical
powercircuit.
ContactInvacareforanyadditionalEMCinformationorEMCtables
forthedeviceenvironment.
12
1145804-E
Set-Up
3Set-Up
3.1Unpacking
PARTNUMBER
DESCRIPTION
1116440CompressorAssembly
1102315PackagingAssembly
1105177SupplyLineAssembly
1131308PowerCord
1145804UserManual(notshown)
1.Checkforanyobviousdamagetothecartonoritscontents.If
damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallofthecomponentsfromthecarton.
UnlessthecompressoristobeusedIMMEDIATELY,leave
thecompressorinitspackagingforstorageuntiluseofthe
compressorisrequired.
3.2CompressorInitialSetupChecklist
DealersONLYReferto2.3DealerInformation,page11
beforeperformingthisprocedure.
UsersONLYContactyourInvacaredealerorInvacare
forassistancewiththesetupprocedureandtoreport
unexpectedoperationorevents.
BeforeoperatingtheHomeFillIICompressorforthefirsttime,
completethefollowingchecklist:
qChoosealocationforthecompressor.
qTransportthecompressortothedesiredlocation.
qConnectthepowercordtothecompressor.
qConnectthecompressortotheconcentrator.
1145804-E13
Invacare®HomeFill®II
3.3ChoosingaLocation
WARNING!
NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor
atmosphere.
Keeptheoxygentubing,cord,andunitawayfrom
heatedorhotsurfaces,includingspaceheaters,
blankets,stovesandsimilarelectricalappliances.
Thecompressorshouldbeatleastthreeinchesawayfromwalls,
draperiesorfurnituretoensuresufficientairflowandclearancefor
thepowercord.
DONOTuseinacloset.
Thecompressorshouldsitonahardflatstablesurface,suchasa
tableortheReadyRack.
TheHomeFillReadyRackissoldseparatelyandusedto
positiontheHomeFillcompressorontopofaPerfecto
2
oxygenconcentrator.RefertotheHomeFillReady
Rackinstructionsheetpartnumber1148363,formore
information.
3.4TransportingtheCompressor
WARNING!
NEVERtransportacompressorwithacylinder
connectedtoorstoredonthecompressor,otherwise
injuryordamagecanoccur.
CAUTION!
Whentransportingthecompressor,becarefulnot
tobumpthecompressorortheconnectorsinto
obstacles.Otherwise,damagetothecompressormay
occur.
Therearetwomethodstosafelytransportthecompressor:
1.Onehand-GraspthehandgripAatthebottomrearofthe
compressor.
2.Twohands-GraspthehandgripsBonthesidesofcompressor.
14
1145804-E
Set-Up
3.5InstallingthePowerCord
1.EnsurethecompressorAisonahardflatstablesurface,suchas
atableorReadyRack.
2.MakesurethepowerswitchisintheOff(O)position.
3.Plugthepowercord(Invacaresupplied)intotheelectrical
connectorBinthebackofthecompressor.
4.PlugtheoppositeendoftheInvacaresuppliedpowercordD
intothewalloutletC.
5.Leaveaminimumof3”(7.6cm)clearancetothepowercord
endsfordisconnectionaccess.
Thepowerinletactsasthemainsdisconnectiondevice
toremoveallpowertotheunitwhenthelinecordis
disconnected.
3.6ConnectingtheCompressortothe
Concentrator
1.InsertoneendBoftheinterconnecthoseA(InvacareSupplied)
intotheoutletfittingBonthebackoftheconcentratorC.
2.Inserttheoppositeendoftheinterconnecthoseintotheinlet
fittingDonthesideofthecompressorE.
ThemetaltabsGontheconcentratoroutletfitting
andthecompressorinletfittingFwillpopoutwithan
audible“click”whentheendoftheinterconnecthose
isproperlyinstalled.
3.Loopanyexcessinterconnecthoseandsecuretothebackof
theconcentratorwiththefasteningstraps.
1145804-E15
Invacare®HomeFill®II
4Usage
4.1CompressorOperationChecklist
EachtimetheHomeFillIICompressorisusedtofillacylinder,
completethefollowingchecklist:
qEnsuretheconcentratorhasbeenOnforatleast30minutes.
RefertotheconcentratorUserManual.
qPerformtheprefillinspectiononthecylinder.
qConnectthecylindertothecompressor.
qPushthecompressorpowerswitchtotheOn(
)position.
qExaminetheindicatorlightsonthecontrolpanel.
qDisconnectandremovethefullcylinder.
qPushthecompressorpowerswitchtotheOff(O)position.
qIffillinganothercylinder,repeatthischecklist.
4.2CylinderPrefillInspection
WARNING!
ONLYusecylindersthathavetheInvacareHomeFillIIconnectionandthecylinderlabel(shownforreference).
AllcylindersMUSTbeinspectedbeforeattemptingtofill.Otherwise,injuryordamagemayoccur.
161145804-E
Usage
4.3TheHydrostaticTestingDate
AluminumcylindersMUSTundergohydrostatictestingeveryfive
years.Contactyourdealerformoreinformation.
WARNING!
DONOTfillcylindersthathavenotbeentestedin
thepastfiveyears.Otherwise,seriouspersonalinjury
mayresult.Contactyourdealerforreplacement.
4.4ExternalExamination
1.Examinetheoutsideofthecylinderforthefollowingconditions,
andreplacethecylinderiftheyexist:
Dentsordings
Arcburns
Oilorgrease
Anyothersignsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse.
2.Examinethecylinderforevidenceoffireorthermaldamage.
Evidenceincludescharringorblisteringofthepaint,orother
protectivecoatingorheatsensitiveindicator.Iffireorthermal
damageisfound,replacethecylinder.
3.Inspectthecylinder/regulatorassemblyforthefollowing,and
replaceiffound:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorfiredamage
4.5Connecting/DisconnectingtheCylinder
to/fromtheCompressor
4.5.1ConnectingtheCylindertotheCompressor
WARNING!
RiskofInjuryorDamage
NEVERusetoolsofanykindtoconnect/disconnect
thecylinderAandthecompressorB.Otherwise,
severeinjuryand/ordamagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,
injuryordamagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
1.ExaminethecylindergaugeC.Ifthecylinderpressureisless
than1500psig(10342kPa),proceedtostep2.
Ifthecylinderpressureisgreaterthan1500psig(10342
kPa),DoNOTattempttotopoffthiscylinder.Itmay
notfill.
2.SetregulatorflowdialDonthecylinderAtoOff.
1145804-E17
Invacare®HomeFill®II
3.Removethecylinderfillportcover
EandconnectorfillportcoverF.
CAUTION!
DONOTconnecttheoutlet(cannula)G
connectiontothecompressor,otherwisethe
conservingcylinderwillnotworkproperly.
4.MomentarilypushDOWNontheouterring(sleeve)Hofthe
connectorfillportIuntilGREENdotsJarevisibletoreset
theconnector.
Iftheouterring(sleeve)isintheUPposition(GREEN
dotsnotvisible),theconnectorfillportwillnotbe
abletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
5.Graspthecylinder/regulatorKassemblyintheareabehindthe
cylindergaugeL.
6.PositionthecylinderinthecompressorcradleM.
7.AlignthecylinderfillportNwiththeconnectorfillportI.
8.PullUPontheouterring(sleeve)oftheconnectorfillportwhile
pushingDOWNonthecylinder/regulatorassemblytocouple
thecylinderfillportintotheconnectorfillport.
Thecylinderisproperlyconnectedwhenanaudible
“click”isheard.
DisconnectingtheCylinderfromtheCompressor
WARNING!
NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe
injuryand/ordamagemayoccur.
1.Whilestandinginfrontoftheunit,pressthecompressorpower
switchtotheOff(O)Oposition.
2.Graspthecylinder/regulatorassemblyKintheareabehindthe
cylindergaugeL.
3.Withtheotherhand,grasptheouterring(sleeve)Hofthe
connectorfillportIandpushDOWN.
181145804-E
Usage
4.Liftuponthecylinder/regulatorassemblytoremovefromthe
connectorfillport.
5.Whenthecylinderfillportisdisconnectedfromtheconnector
fillport,releasetheouterring(sleeve)oftheconnectorfillport
andusetwohandstoremovethecylinderfromthecompressor
cradle.
WARNING!
Thefillportcoverontheconnectorfillportand
thecylinderfillportMUSTbereplacedafter
fillingandwhenevernotinuse.Ifeitherofthe
fillportcoversaremissingcontactyourhealthcare
providerbeforeusing.
6.Placethefillportcoversontothecon-
nectorfillportFandcylinderfillportE.
7.Examinethecylindergaugetoensurecylinderisfull(needle
pointingintoGREENarea).Ifthecylinderisnotfull,perform
thefollowing:
Repeat4.5.1ConnectingtheCylindertotheCompressor,
page17
.
Ifcylinderisnotfullaftersecondattempttofill,contact
theprovider.
4.6TurningtheCompressorOnandOff
1145804-E19
Invacare®HomeFill®II
1.MakesuretheconcentratorisOn.Refertotheconcentrator
UserManual.
Themaximumoxygenconcentratorflowratefor
patientuseislistedinthetablebelow.Otherwise,the
oxygenleveltothecompressorwillbe<90%andthe
compressorwillnotfillthecylinder.
InvacareOxygen
Concentrator
MaximumOutputFlow
forPatientUse
Platinum5
2.5L/min
Platinum95L/min
Perfecto
2
3L/min
Patientswithahigherprescribedflowshouldnotrefilla
cylinderwhilereceivingsupplementaloxygentreatment
fromtheoxygenconcentrator.
2.Settheconcentratorflowratetothepatientto2.5L/min.orless
forPlatinum5and5L/min.forPlatinum9oxygenconcentrator.
RefertotheconcentratorUserManual.
3.Makesurethecylinderisconnectedtothecompressor.Refer
to4.5.1ConnectingtheCylindertotheCompressor,page17.
4.PushthepowerswitchBonthecontrolpanelAtotheOn
()position.
5.Examinethecontrolpanel.Thefollowingsequenceofevents
shouldoccur:
a.0-3minutes-TheO
2
belownormal(YELLOW)lightCwill
belitwhilethecompressorwarmsup.
b.Afterthreeminutes-Thefilling(green)lightDwillbeon
whilethecylinderisfilling.
c.Thefull(green)lightEwillbeonwhenthecylinderis
finishedfilling.Proceedtostep6.
TheO
2
belownormal(YELLOW)lightmaycomeon
iftheoxygenlevelfromtheconcentratorhasnotyet
reachedordropsbelow90%orhigher.Thecompressor
willbeginorresumefillingwhentheoxygenlevel
reaches90%.Referto7.2CylinderFillTimes,page27
if
lightdoesnotgoOffwithin10minutes.
Referto7.2CylinderFillTimes,page27
iftheAttention
(red)lightFandaudiblealarmareon.
6.Removethefullcylinder.RefertoDisconnectingtheCylinder
fromtheCompressor,page18.
7.PushthepowerswitchtotheOff(O)position.
Forcylinderoperationanduse,refertothespecificuser
manualforthecylinder/regulatorassembly.
201145804-E
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180

Invacare IOH200AW Benutzerhandbuch

Typ
Benutzerhandbuch