Invacare Birdie EVO Benutzerhandbuch

Typ
Benutzerhandbuch

Dieses Handbuch eignet sich auch für

InvacarBirdie®EVO
Birdie®EVO,Birdie®EVOCOMPACT,Birdie®EVOPLUS,
Birdie®EVOXPLUS
enMobilePatientLift
UserManual...................................3
deMobilerPatientenlifter
Gebrauchsanweisung...........................35
esGrúadeelevación
Manualdelusuario.............................71
frLève-personnemobile
Manueld'utilisation............................103
itSollevatoremobile
Manualed’uso.................................137
nlMobielepatiëntenlift
Gebruiksaanwijzing.............................169
ptElevadordetransferênciamóvel
Manualdeutilização...........................201
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsaved
forfuturereference.
©2019InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibitedwithoutpriorwritten
permissionfromInvacare.Trademarksareidentiedbyand®.AlltrademarksareownedbyorlicensedtoInvacare
Corporationoritssubsidiariesunlessotherwisenoted.
AlleRechtevorbehalten.DieWeiterveröffentlichung,VervielfältigungoderÄnderungimGanzenoderinTeilenohne
vorherigeschriftlicheGenehmigungvonInvacareistuntersagt.Markensinddurchund®gekennzeichnet.Soweit
nichtandersangegebensindalleMarkenEigentumderInvacareCorporationbzw.dererTochtergesellschaftenoder
werdenvondieseninLizenzgenutzt.
Todoslosderechosreservados.Quedaprohibidovolverapublicar,copiaromodicarelpresentedocumento,en
parteoporcompleto,sinelprevioconsentimientoporescritodeInvacare.Lasmarcascomercialesestánidenticadas
mediantey®.Amenosqueseestipulelocontrario,todaslasmarcascomercialessonpropiedaddeInvacare
Corporationodesusliales,obienInvacareCorporationosuslialescuentanconlascorrespondienteslicencias.
Tousdroitsréservés.Larepublication,laduplicationoulamodicationdetoutoupartieduprésentdocumentest
interditesansl'accordécritpréalabled'Invacare.Lesmarquescommercialessontidentiéesparet®.Toutes
lesmarquescommercialessontdétenuesparoucédéessouslicenceàInvacareCorporationousesliales,sauf
stipulationcontraire.
Tuttiidirittiriservati.Èproibitalariproduzione,laduplicazioneolamodicaparzialeocompleta,salvoprevia
autorizzazionescrittadapartediInvacare.Imarchisonocontrassegnatidae®.Tuttiimarchisonodiproprieo
licenzadiInvacareCorporationodisueafliate,salvoindicazionecontraria.
Allerechtenvoorbehouden.Herpublicatie,duplicatieofgeheleofgedeeltelijkewijzigingisverbodenzonder
voorafgaandeschriftelijketoestemmingvanInvacare.Handelsmerkenzijnteherkennenaanen®.Allehandelsmerken
zijneigendomvanofgelicentieerdaanInvacareCorporationofhaardochterondernemingen,tenzijandersaangegeven.
Todososdireitosreservados.Arepublicação,duplicaçãooumodicaçãototalouparcialestáinterditasema
autorizaçãopréviaporescritodaInvacare.Asmarcascomerciaissãoidenticadaspelossímbolose®.Todasas
marcascomerciaissãopropriedadedaouestãolicenciadasàInvacareCorporationouàssuassubsidiárias,exceto
quandoapresentadainformaçãoemcontrário.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsaved
forfuturereference.
1General.........................................4
1.1Introduction..................................4
1.1.1Symbolsinthisdocument.....................4
1.2ServiceLife...................................4
1.2.1Additionalinformation........................4
1.3LimitationofLiability............................4
1.4WarrantyInformation...........................4
1.5Compliance...................................4
1.5.1Product-specicstandards.....................4
2Safety..........................................5
2.1Generalsafetyinformation.......................5
2.1.1Pinchpoints................................5
2.2Safetyinformationonaccessories..................6
2.3Safetyinformationonelectromagneticinterference.....6
2.4Labelsandsymbolsontheproduct.................6
2.4.1Labellocation..............................6
2.4.2Identicationlabel...........................7
2.4.3Othersymbols..............................7
3ProductOverview.................................8
3.1Intendeduse..................................8
3.2Mainpartsofthelift............................8
3.3Accessories...................................8
4Setup...........................................9
4.1Generalsafetyinformation.......................9
4.2Scopeofdelivery..............................9
4.3Installingthemast..............................9
4.3.1Unfoldingthemast..........................9
4.3.2Assemblingthemasttothebase................10
4.4Unfoldingthespreaderbar.......................10
4.5Installingtheactuatortotheboom.................11
4.6Installingtheleverformanuallegspreader...........11
4.7Resettingtheserviceindicator.....................11
5Usage..........................................13
5.1Generalsafetyinformation.......................13
5.2Locking/Unlockingtherearcastors.................13
5.3Raising/LoweringanElectricLift...................13
5.4Closing/OpeningLegs...........................13
5.4.1Closing/OpeningElectricLegs...................13
5.4.2Closing/Openinglegsmanually..................13
5.5Replacingthespreaderbar.......................13
5.6Emergencyfunctions............................14
5.6.1Performinganemergencystop..................14
5.6.2Activatinganemergencylowering(CBJHome
controlunit)...............................14
5.6.3Activatinganemergencylowering(CBJCare,CBJ1,
CBJ2controlunit)...........................14
5.6.4Activatinganemergencylifting(CBJCare,CBJ1
controlunit)...............................14
5.6.5Activatingamechanicalemergencylowering.......14
5.7Chargingthebattery............................15
5.7.1CBJHomecontrolunit........................15
5.7.2CBJCare,CBJ1,CBJ2controlunit................15
5.7.3Optionalhandcontrol........................16
5.7.4Optionalbatterycharger......................16
6PatientTransfer...................................18
6.1Generalsafetyinformation.......................18
6.2Preparingtolift................................18
6.2.1Attachingtheslingtothelift...................19
6.3Transferringapatientfromabed..................19
6.4Transferringapatienttoabed....................20
6.5Transferringapatientfromawheelchair.............21
6.6Transferringapatienttoawheelchair...............21
6.7Transferringapatienttoandfromacommode........21
6.8Liftingapatientfromtheoor....................22
7TransportationandStorage..........................24
7.1Generalinformation............................24
7.2Disassemblingthemastfromthebase...............24
7.3Foldingthemast...............................24
7.4Mastextensionforuprightpositioning...............24
8Maintenance.....................................25
8.1Generalmaintenanceinformation..................25
8.2Dailyinspections...............................25
8.3CleaningandDisinfection........................25
8.3.1Generalsafetyinformation.....................25
8.3.2CleaningIntervals............................25
8.3.3Cleaninginstructions.........................25
8.3.4Disinfectioninstructions.......................26
8.4Serviceinterval................................26
8.4.1LOLERStatement............................26
9AfterUse........................................27
9.1Disposal.....................................27
9.2Reconditioning................................27
10Troubleshooting..................................28
10.1Identifyingfaultsandpossiblesolutions.............28
11TechnicalData...................................29
11.1Maximumsafeworkingload.....................29
11.2Dimensionsandweights........................29
11.3Electricalsystem..............................30
11.4Environmentalconditions........................31
11.5Materials....................................31
11.6Operatingforcesofcontrols......................31
12Electromagneticcompatibility(EMC)..................32
12.1GeneralEMCinformation.......................32
12.2Electromagneticemission........................32
12.3ElectromagneticImmunity.......................32
InvacarBirdie®EVO
1General
1.1Introduction
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.Toensuresafetywhenusingthe
product,readtheusermanualcarefullyandfollowthe
safetyinstructions.
Notethattheremaybesectionsinthisdocument,whichare
notrelevanttoyourproduct,sincethisdocumentapplies
toallavailablemodels(onthedateofprinting).Ifnot
otherwisestated,eachsectioninthisdocumentreferstoall
modelsoftheproduct.
Themodelsandcongurationsavailableinyourcountrycan
befoundinthecountry-specicsalesdocuments.
Invacarereservestherighttoalterproductspecications
withoutfurthernotice.
Beforereadingthisdocument,makesureyouhavethe
latestversion.YoundthelatestversionasaPDFonthe
Invacarewebsite.
Ifyoundthatthefontsizeintheprinteddocumentis
difculttoread,youcandownloadthePDFversionfromthe
website.ThePDFcanthenbescaledonscreentoafont
sizethatismorecomfortableforyou.
Formoreinformationabouttheproduct,forexampleproduct
safetynoticesandproductrecalls,contactyourInvacare
distributor.Seeaddressesattheendofthisdocument.
Incaseofaseriousincidentwiththeproduct,youshould
informthemanufacturerandthecompetentauthorityin
yourcountry.
1.1.1Symbolsinthisdocument
Symbolsandsignalwordsareusedinthisdocumentand
applytohazardsorunsafepracticeswhichcouldresultin
personalinjuryorpropertydamage.Seetheinformation
belowfordenitionsofthesignalwords.
WARNING
Indicatesahazardoussituationthatcouldresult
inseriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresult
inminororslightinjuryifitisnotavoided.
IMPORTANT
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
TipsandRecommendations
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
1.2ServiceLife
Theexpectedservicelifeofthisproductiseightyearswhen
useddailyandinaccordancewiththesafetyinstructions,
maintenanceintervalsandcorrectuse,statedinthismanual.
Theeffectiveservicelifecanvaryaccordingtofrequencyand
intensityofuse.
1.2.1Additionalinformation
Theexpectedservicelifeisbasedonanestimatedaverage
of4liftingcyclesperday.
1.3LimitationofLiability
Invacareacceptsnoliabilityfordamagearisingfrom:
Non-compliancewiththeusermanual
Incorrectuse
Naturalwearandtear
Incorrectassemblyorset-upbythepurchaserorathird
party
Technicalmodications
Unauthorisedmodicationsand/oruseofunsuitable
spareparts
1.4WarrantyInformation
Weprovideamanufacturerswarrantyfortheproduct
inaccordancewithourGeneralTermsandConditionsof
Businessintherespectivecountries.
Warrantyclaimscanonlybemadethroughtheprovider
fromwhomtheproductwasobtained.
1.5Compliance
Qualityisfundamentaltothecompany’soperation,working
withinthedisciplinesofISO13485.
ThisproductfeaturestheCEmark,incompliancewiththe
MedicalDeviceRegulation2017/745ClassI.Thelaunchdate
ofthisproductisstatedintheCEdeclarationofconformity.
Wearecontinuouslyworkingtowardsensuringthatthe
company’simpactontheenvironment,locallyandglobally,
isreducedtoaminimum.
WeonlyuseREACHcompliantmaterialsandcomponents.
WecomplywiththecurrentenvironmentallegislationsWEEE
andRoHS.
1.5.1Product-specicstandards
TheproducthasbeentestedandconformstoISO10535
(Hoistsforthetransferofdisabledpersons)andallrelated
standards.
Forfurtherinformationaboutlocalstandardsandregulations,
contactyourlocalInvacarerepresentative.Seeaddresses
attheendofthisdocument.
41650023-D
Safety
2Safety
2.1Generalsafetyinformation
Thissectionofthemanualcontainsgeneralsafety
informationaboutyourproduct.Forspecicsafety
information,refertotheappropriatesectionofthemanual
andprocedureswithinthatsection.
WARNING!
Riskofinjuryordamage
Donotusethisproductoranyavailableoptional
equipmentwithoutrstcompletelyreading
andunderstandingtheseinstructionsandany
additionalinstructionalmaterialsuchasuser
manualsorinstructionsheetssuppliedwith
thisproductoroptionalequipment.Ifyouare
unabletounderstandthewarnings,cautionsor
instructions,contactahealthcareprofessional,
Invacareproviderorqualiedtechnicianbefore
attemptingtousethisproduct.
Donotmakeanyunauthorizedalterationsor
modicationstotheproduct.
WARNING!
Themaximumsafeworkingloadmustnotbe
exceeded
Donotexceedthemaximumsafeworkingload
ofthisproductorusedaccessorieslikeslings,
spreaderbarsetc.Seedocumentationorthe
labelingforthestatedmaximumsafeworking
load.
Thecomponentwiththelowestloadlimit
determinesthemaximumsafeworkingloadof
theentiresystem.
WARNING!
Riskofinjuryordamage
Improperuseofthisproductmaycauseinjury
ordamage.
Donotattemptanytransferwithoutapproval
ofthepatientshealthcareprofessional.
Readtheinstructionsinthisusermanualand
observetrainedpersonnelperformingtransfer
procedures.Thenpracticetransfersunder
supervisionandwithacapablepersonacting
asapatient.
Specialcaremustbetakenwithpeoplewith
disabilitieswhocannotcooperatewhilebeing
transferred.
Donotusetheliftasatransportdevice.It
isintendedtotransferanindividualfromone
restingsurfacetoanother.
WARNING!
Riskofinjuryordamage
Inappropriatehandlingofcablescancause
electricalshockandproductfailure.
Donotkink,shearorotherwisedamagethe
cablesoftheproduct.
Makesurethatnocablesarejammedor
damaged,whentheproductisused.
Makesurethewiringiscorrectandconnections
areproper.
Donotuseunauthorizedequipment.
WARNING!
Riskofinjuryordamage
Excessivemoisturewilldamagetheproductand
maycauseelectricalshock.
Thepatientliftcanbeusedinabathorshower
areabutmustNOTbeusedundertheshower.
Thepatientmustbetransferredtoashower
chairoruseothermeansforshowering.
Ifthepatientliftisusedinamoisture
environment,ensurethatthepatientliftis
wipedcleanofanymoistureafteruse.
Donotplugorunplugthepowercableina
moistureenvironmentorwithwethands.
Donotstoretheproductinadampareaor
dampcondition.
Periodicallyinspectallcomponentsofthe
productforsignsofcorrosionordamage.
Replacepartsthatarecorrodedordamaged.
See11.4Environmentalconditions,page31.
WARNING!
Riskofinjuryordamage
Ignitionsourcescancauseburnsorre.
Patienttransfermustbeperformedwitha
safetyspacebetweentheliftandpossible
ignitionsources(heater,stove,replace,etc.)
Thepatientandassistantsmustnotsmoke
duringtransfer.
Theslingmustnotbeplacedoverheatsources
(heater,stove,replace,etc.)
WARNING!
Riskofinjuryordamage
Toavoidinjuryordamagewhenoperatingthe
product:
Closesupervisionisnecessarywhentheproduct
isusednearchildrenorpets.
Donotletchildrenplaywiththeproduct.
CAUTION!
Riskofinjuryordamage
Theproductcangethotwhenexposedtosunlight
orotherheatsources.
Donotexposetheproducttodirectsunlightfor
prolongedperiods.
Keeptheproductawayfromheatsources.
IMPORTANT!
Accumulationoflint,dustandotherdirtcan
impairtheproduct.
Keeptheproductclean.
2.1.1Pinchpoints
1650023-D5
InvacarBirdie®EVO
WARNING!
RiskofInjury
Pinchpointsarepresentinseverallocationson
theliftandngerscouldbepinched.
Alwayskeephandsandngersclearofmoving
parts.
2.2Safetyinformationonaccessories
CAUTION!
Riskofinjury
Non-originalorincorrectaccessoriesmayaffect
thefunctionandsafetyofthisproduct.
Duetoregionaldifferences,refertoyour
localInvacarewebsiteorcatalogforavailable
accessoriesorcontactyourInvacareprovider.
Seethemanualdeliveredwiththeaccessory
forfurtherinformationandinstructions.
Useonlyoriginalaccessoriesfortheproduct
inuse.Undercertaincircumstances,theuse
slingsfromothermanufacturersispossible.See
additionalinformationinthissection.
Ariskassessmentisalwaystobecarriedoutby
aprofessionalpriortoissuingliftingequipment.
ItisimportantthattheTask,Individual,Load,
EnvironmentandEquipmentareconsideredin
theriskassessment.
Alwayschoosetheslingdesignandsize
accordingtothepatient’sweight,sizeand
physicalabilitywhilstconsideringthetypeof
transfertobecarriedout.
CAUTION!
Compatibilityofslingswithattachmentsystem
Invacareusesacommonattachmentsystem
basedonhooksandloops.Loopsontheslings
areattachedtohooksonspreaderbars(coat
hangerstyle).Therefore,suitableslingsofother
manufacturerscanbeusedonthisliftaswell.
Useonlyslingswithloopattachmentsthatare
suitableforspreaderbars(coathangerstyle)
withhooks.
Donotuseslingsdesignedfor"Keyholeor
ClipAttachmentSystems"or"TiltingFrame
Systems".
2.3Safetyinformationon
electromagneticinterference
WARNING!
Riskofmalfunctionduetoelectromagnetic
interference
Electromagneticinterferencebetweenthis
productandotherelectricalequipmentcanoccur
anddisturbtheelectricaladjustmentfunctions
ofthisproduct.Toprevent,reduceoreliminate
suchelectromagneticinterference:
Onlyuseoriginalcables,accessoriesandspare
parts,tonotincreaseelectromagneticemission
orreduceelectromagneticimmunityofthis
product.
Donotuseportableradiofrequency(RF)
communicationsequipmentcloserthan30cm
toanypartofthisproduct(includingcables).
Donotusethisproductnearactive
high-frequencysurgicalequipmentandthe
RFshieldedroomofasystemformagnetic
resonanceimaging,wheretheintensityof
electromagneticdisturbancesishigh.
Ifdisturbancesoccur,increasethedistance
betweenthisproductandtheotherequipment
orswitchitoff.
Refertothedetailedinformationandfollow
theguidanceinchapter12Electromagnetic
compatibility(EMC),page32.
WARNING!
Riskofmalfunction
Electromagneticinterferencemaycauseimproper
operation.
Donotusethisproductadjacenttoorstacked
withotherelectricalequipment.Ifsuchuseis
necessarythisproductandtheotherequipment
mustbecloselyobservedtoverifythatthey
areoperatingnormally.
2.4Labelsandsymbolsontheproduct
2.4.1Labellocation
AIdenticationLabel
BReadusermanuallabel
CWarninglabelonactuator
DWarrantyvoidlabelonCBJ1andCBJ2control
unitonly
EEnvironmentalconditionsandtotalmaximum
weightlabel
FSafeworkingloadofspreaderbar
GNameandsafeworkingloadoflifttextdepending
onmodel
HSafeworkingloadoflifttextdependingonmodel
(xedmastonly)
61650023-D
Safety
2.4.2Identicationlabel
Product Name
YYYY-MM
= xxx kg
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio, Portugal
XXXXXX XXXXX_XX
XXXXXXXXXXXXX
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
ISO 10535
Theidenticationlabelcontainsthemainproduct
information,includingtechnicaldata.
Symbols
Serialnumber
Referencenumber
Manufacturer
Dateofmanufacture
Max.safeworkingload
ClassIIequipment
TypeBappliedpart
WEEEconform
EuropeanConformity
Medicaldevice
Abbreviationsfortechnicaldata:
Iin=IncomingCurrent
Uin=IncomingVoltage
Int.=Intermittence
AC=AlternatingCurrent
Max=maximum
min=minute
Formoreinformationabouttechnicaldata,see11Technical
Data,page29.
2.4.3Othersymbols
RefertoUserManual
Totalweightoftheproductwiththe
maximumsafeworkingloadapplied
Donotusetheactuatorasapushbar
Donotremove
Warrantyvoidifremovedorbroken
Temperaturelimit
Humiditylimitation
Atmosphericpressurelimitation
Transportandstorageconditions
Operationconditions
1650023-D7
InvacarBirdie®EVO
3ProductOverview
3.1Intendeduse
Themobilepatientliftisabattery-poweredtransferdevice
andisintendedtotransferandpositionanindividualfrom
onerestingsurfacetoanother.Forexample:
Betweenthebedandawheelchair
Toandfromthetoilet
Loweringandraisingpatientsto/fromtheoor
Themaximumsafeworkingloadisstatedin11Technical
Data,page29
Themobilepatientliftisdesignedtobeusedindoorsona
levelsurface,inhospitals,nursingfacilitiesanddomestic
areas.Themobilepatientliftcanbeturned(rotated)in
placefortransferswithlimitedoorspace.
Ahealthcareprofessionalorprivatepersonwhohasreceived
propertrainingistheintendedoperatorofthisproduct.
Indications
Themobilepatientliftisindicatedforcompletelyorpartially
immobilepatients,whocannotbetransferredwithother
typesofliftsortransferaids.
Allpositionchangesarepossiblewithoutassistanceofthe
patient.
Therearenoknowncontraindicationsforthisproduct.
3.2Mainpartsofthelift
ABoom
BSpreaderbarwithorwithoutSMARTLOCK™
CHookforsling
DPushbar
ELiftingactuator
FMastfoldableorxed
GBasewithfootpedalformanuallegspreader
HFrontCastors
ILeg
JBasewithactuatorsforelectricallegspreader
withorwithoutactuatorcovers
KRearcastorswithbrake
LHandcontrol
MEmergencyStop
NCBJHomecontrolunitwithintegratedbattery
OCBJCare,CBJ1orCBJ2controlunitwithdetachable
battery
3.3Accessories
Duetoregionaldifferences,refertoyourlocal
Invacarewebsiteorcatalogfordetailsaboutavailable
accessoriesorcontactyourInvacareprovider.
4–pointspreaderbar(coathangerstyle)450,500or
550mmwide
2–pointspreaderbar(coathangerstyle)350,450or
550mmwide
Scaletobemountedwiththespreaderbar
Leverformanuallegspreader
Wallchargerforadetachablebattery
Extrabattery
Protectiveshroudsforlegs
Paddingforspreaderbar
Slingmodelswithloopattachments,suitableforspreader
bars(coathangerstyle)withhooks:
Fullbodysupportslings-withoutheadsupport
Fullbodysupportslings-withheadsupport
Slingsfordress/toileting-withorwithoutheadsupport
Slingsforamputee
81650023-D
Setup
4Setup
4.1Generalsafetyinformation
WARNING!
Checkallpartsforshippingdamagebeforeuse.
Incaseofdamage,donotusetheequipment.
ContactyourInvacareproviderforfurther
instructions.
WARNING!
RiskofInjury
Improperassemblymaycauseinjuryordamage.
UseonlyInvacarepartsintheassemblyofthis
patientlift.
Aftereachassembly,checkthatallttings
areproperlytightenedandallpartshavethe
correctfunction
Donotovertightenthemountinghardware.
Thismaydamagethemountingbrackets.
Therearenotoolsrequiredtoassemblethepatient
lift.
Ifthereareanyissuesorquestionsduringassembly,
contactyourInvacareprovider.
4.2Scopeofdelivery
Theitemsincludedinyourpackagedependonthemodels
andcongurationsavailableinyourcountry.See1.1
Introduction,page4
ALiftincl.base,mast,boom,spreaderbar,
controlunitandactuators
withfoldablemast(1piece)
withxedmast(2pieces)baseasa
separatepart
BHandcontrol(1piece)
Mainscable(1piece)
UserManual(1piece)
Battery(1piece)*CBJCare,CBJ1andCBJ2
only
Leverformanuallegspreader(1piece)*
Sling(1piece)*
*Dependsonmodeland/orconguration
Iftheliftisdeliveredwithasling,refertotheuser
manualoftheslingforuse,application,maintenance
andwashinginstructions.
4.3Installingthemast
4.3.1Unfoldingthemast
(foldablemastonly)
WARNING!
Themastmaybefoldedforstorageor
transporting.Eachtimethemastisfolded,the
mastMUSTbeproperlysecuredtothebase
assembly.
Checkallpartsforvisibledefectsordamage
beforeassembly.Incaseofanydamage,do
notusetheproductandcontactyourInvacare
provider.
Makesuretheemergencystopisactivated
beforeassemblyordisassembly.
Takecarewhenliftingcomponentsduring
assembly.Somepartsareheavy.Always
remembertoadoptthecorrectliftingposition.
Performunpackingandassemblyoperationatoorlevel.
1650023-D9
InvacarBirdie®EVO
1.LockbothrearcastorsB.RemovethelockingpinA.
2.RaisethemastassemblyCtoanuprightpositionby
steppingwithonefootonthelegDandpullingthepush
barEupwardsuntilthesafetycatchFsnapsinplace.
3.
ReinstallthelockingpinAthroughthemastGand
baseH.Ensurethatthelockingpiniscorrectlyinserted.
4.3.2Assemblingthemasttothebase
(xedmastonly)
WARNING!
RiskofinjuryorDamage
Themaximumsafeworkingloadstatedonthe
boomandonthebasemustbethesame.
Alwayscomparethemaximumsafeworking
loadvaluesstatedonthebaseandonthe
boom.
1.
LockbothrearcastorsBandrotatethehandscrews
anti-clockwiseAtoremovethemfromthebase
2.
Withtheboompointinginforwarddirection,lowerthe
mastCintothesupportofbaseDandreinstallthe
handscrewsAtoxthemastCtothebase.
4.4Unfoldingthespreaderbar
1.
OpenthefastenerAandremovethebandBfromthe
clampingbuckle.
101650023-D
Setup
2.
PushandholdtheboomCtowardsthemastDand
folddownthespreaderbarholderE.
3.
FolddownthespreaderbarFandfoldupthespreader
barholderE.
4.
WrapthebandBaroundthebackofthemast,insertit
intotheclampingbuckleandclosethefastenerA.
4.5Installingtheactuatortotheboom
1.ReleasetheD-clipandremovethepinAfromactuator
B.
2.RemovetheactuatorBfromthebracketonthemast
andtittotheboommountingbracketC.
3.
AligntheholesandreinstallthepinA.Makesurethat
thepiniscompletelyinsertedandsecureitwiththe
D-clip.
IMPORTANT
Duetohigherloads,Birdie®EVOXPLUSrequires
alongerpin.
OnlyuseBirdie®EVOXPLUSwiththelonger
pin,indicatedbyablackD-Clip.
4.6Installingtheleverformanualleg
spreader
2x13mmwrench
1.InserttheboltAfrombelowintothelowerholeofthe
bracketBandxitwiththenutC.
2.InsertthethreadedendofleverDintotheupperhole
ofbracketBontothexedboltA.
3.TurntheleverDclockwisetoscrewitontothebolt.
4.7Resettingtheserviceindicator
(CBJCarecontrolunitonly)
IMPORTANT!
Iftheservicelightisashingyellowduringdaily
usageorafterareassembly,theliftrequires
service.
ContactyourInvacareproviderforservice.
Aftertheinitialassemblyofthelift,theserviceindicator
needstoberesetbeforeusingthelift.
Toperformaninitialresetoftheserviceindicator:
1650023-D11
InvacarBirdie®EVO
1.Locatethehandcontrol.
2.PressandholdtheUPbuttonandtheDOWNbuttonat
thesametimeforveseconds.
3.Youwillhearasoundwhentheserviceindicatorhas
beenreset.
121650023-D
Usage
5Usage
5.1Generalsafetyinformation
WARNING!
Riskofinjuryordamage
Beforeusingtheliftwithapatient,refertothe
followingsafetyinformationandinstructions:
2Safety,page5
6PatientTransfer,page18
5.2Locking/Unlockingtherearcastors
WARNING!
RiskofInjuryorDamage
Theliftcouldtipandendangerthepatientand
assistants.
Invacarerecommendsthattherearcastors
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedorany
stationaryobject.
Invacarerecommendslockingtherearcastors
oftheliftonlywhenpositioningorremoving
theslingfromaroundthepatient.
Tolockthecastor,pushdownpedalAwiththefoot.
Tounlockthecastor,pushuppedalAwiththefoot.
5.3Raising/LoweringanElectricLift
Thehandcontrolisusedtoraiseorlowerthelift.
1.ToraisetheliftPress
andholdtheUPAbutton
toraisetheboomandthe
patient.
2.TolowertheliftPress
andholdtheDOWNB
buttontolowertheboom
andthepatient.
Releasethebuttontostopraisingorloweringthelift.
5.4Closing/OpeningLegs
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegsof
theliftonlyaslongasittakestopositionthe
liftoverthepatientandliftthepatientoffthe
surfaceofthebed.Whenthelegsofthelift
arenolongerunderthebed,returnthelegsof
thelifttothemaximumopenposition.
5.4.1Closing/OpeningElectricLegs
Thehandcontrolisusedtoopenorclosethelegsofthe
base.
1.Toclosethelegs,press
andholdthelegsclosed
buttonA.
2.Toopenthelegs,pressand
holdthelegsopenbutton
B.
Thelegswillstopmovingwhenthebuttonis
released.
5.4.2Closing/Openinglegsmanually
Themanuallegspreaderisoperatedbytwopedals(Aand
B)orbytheleverC.
1.Toopenthelegs,presstherightpedalBwithafoot.
2.Toclosethelegs,presstheleftpedalAwithafoot.
Withthelever:
1.Toopenthelegs,pulltheleverCtotheright.
2.Toclosethelegs,pushtheleverCtotheleft.
5.5Replacingthespreaderbar
(spreaderbarwithSMARTLOCK™only)
WARNING!
Riskofinjury
Useonlyspreaderbarsmadeforthislift.
Makesurethespreaderbarissuitableforthe
patientandtheactualliftortransferrequired.
Checkthatthespreaderbarisrmlyattached
totheboomconnectorandcannotberemoved
withoutpressingthereleasebutton.
1650023-D13
InvacarBirdie®EVO
Removingthespreaderbar
1.2.
1.Pushandholdthereleasebuttonup.
2.Slidethespreaderbaroutinaforwardandslightly
upwarddirection.
Attachingaspreaderbar
1.Slidethespreaderbar
inuntilyouhearan
audibleclick.
2.Checkthatthespreader
barisrmlyattached
totheboomconnector
andcannotberemoved
withoutpressingthe
releasebutton.
5.6Emergencyfunctions
5.6.1Performinganemergencystop
1.PresstheredemergencybuttonAonthecontrolunit
tostoptheboomandpatientfromraisingorlowering.
2.Toreset,rotatetheemergencybuttonclockwise.
5.6.2Activatinganemergencylowering(CBJ
Homecontrolunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing
thecircularswitchforemergencyrelease.
1.LowertheboombypressingandholdingbuttonAat
thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
5.6.3Activatinganemergencylowering(CBJ
Care,CBJ1,CBJ2controlunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing
thecircularswitchforemergencyrelease.Useasharp
object,suchasapenciltopressthebutton.
1.LowertheboombypressingandholdingbuttonAat
thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
5.6.4Activatinganemergencylifting(CBJCare,
CBJ1controlunit)
Ifthehandcontrolfails,theboomcanberaisedbyusing
thecircularswitchforemergencylifting.Useasharpobject,
suchasapenciltopressthebutton.
1.RaisetheboombypressingandholdingbuttonBatthe
frontofthecontrolunit.
2.Stopraisingtheboombyreleasingthebutton.
5.6.5Activatingamechanicalemergency
lowering
Iftheemergencyloweringonthecontrolunitisnotworking,
amechanicalemergencyloweringisavailableasaback-up.
Thismayhappenincaseofpartialortotalpowerfailure,or
ifthebatteryrunsdownduringuse.
WARNING!
RiskofInjuryorDamage
Repeatedmechanicalemergencyloweringleads
toahighincreaseoftheloweringspeed.
Onlyusethemechanicalemergencylowering
asaback-upwhentheemergencyloweringon
thecontrolunitisnotworking.
Donotusethemechanicalemergencylowering
repeatedlyinclosesuccession.
Alwayslowerthepatientontoasupporting
surfacelikeabedorachair.
Afteramechanicalemergencyloweringwas
required,theliftmustbecheckedtoresolve
themalfunctionofthecontrolunit.
141650023-D
Usage
Birdie®EVOandBirdie®EVOPLUS/XPLUS
1.Locatethered
emergencygripA
atthebottomofthe
actuatorpiston.
2.Slowlypullupthered
emergencygripAand
holditinpositionwhen
asafeloweringspeed
isreached.
3.Ifnoloweringoccurs
withafullypulled
emergencygripA,
pushdownonthe
boomatthesame
time.
Theeffectiveloweringspeedisdependingonthepatient
weight.Ifthespeedistoolowortoohigh,itcanbeadapted
tothepatientweight.
Adjustingtheloweringspeed:
1.Locatethescrewinthebottomholeofthered
emergencygripA.
2.Turnthescrewcounter-clockwisetoincreasethespeed.
3.Turnthescrewclockwisetodecreasethespeed.
Birdie®EVOCOMPACT
(optional)
1.Locatetheemergency
releaseknobAatthe
topoftheactuator
piston.
2.Rotatetheemergency
releaseknobA
clockwise.
5.7Chargingthebattery
IMPORTANT!
Makesuretheemergencystopisnotactivated
whilechargingthebattery.
Makesurethatchargingtakesplaceinaroom
withgoodairventilation.
Electricalfunctionsarenotworkingwhenthe
liftisconnectedtopowersupply.
Donotattempttousetheliftifthebattery
housingisdamaged.
Replaceadamagedbatteryhousingbefore
furtheruse.
Donotmovetheliftwithoutunpluggingfrom
thepoweroutlet.
Itisrecommendedtochargethebatterydailytoensure
optimaluseoftheliftandprolongthelifeofthebattery.
Furthermore,itisrecommendedtochargethebattery
beforerstuse.
5.7.1CBJHomecontrolunit
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
1.PlugthepowercordCintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe
batteriesarefullycharged.
TheupperyellowdiodeAwillashduring
charging,andswitchtocontinuouslightwhen
fullycharged.
ThelowergreendiodeBwilllightcontinuously
whilethecontrolunitisconnectedtothe
mains,andlightsupwhenanybuttononthe
handcontrolispressedorwhentheelectric
emergencyloweringisactivated.
2.Disconnectthepowercordfromthepoweroutletafter
thebatteryhasbeenfullycharged.
5.7.2CBJCare,CBJ1,CBJ2controlunit
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
1.PlugthepowercordAintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe
batteriesarefullycharged.
TherightyellowdiodeBwilllightcontinuously
duringcharging,andswitchoffwhenfully
charged.
TheleftgreendiodeCwilllightcontinuously
whilethecontrolunitisconnectedtothemains.
2.DisconnectthepowercordAfromthepoweroutlet
afterthebatteryhasbeenfullycharged.
BatteryIndicator
Thecontrolunitmaybeequippedwithabatteryindicator
D,indicatingtheremainingbatterycapacity.
1650023-D15
InvacarBirdie®EVO
CBJCare
Battery
Indicator
Type
Battery
State
Description
Full
Charge
ThebatteryisOKnoneed
forcharging(100–50%).The
topLEDisGREEN.
Partial
Charge
Thebatteryneedstobe
charged(50–25%).Themiddle
LEDisYELLOW.
Low
Charge
Thebatteryneedstobe
charged(Lessthan25%).
Acousticsignalwhenabutton
ispressed.ThebottomLEDis
YELLOW.
Low
Charge
(LED
ashing)
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
CBJ1andCBJ2withLCD
Battery
Indicator
Type
Battery
State
Description
Full
Charge
(100%)
ThebatteryisOKnoneed
forcharging(100%).
Partial
Charge
(75%)
ThebatteryisOKnoneed
forcharging(75%).
Partial
Charge
(50%)
Thebatteryneedstobe
charged(50%).
Low
Charge
(25%)
Thebatteryneedstobe
charged(25%).Acousticsignal
whenabuttonispressed.
Low
Charge
(0%)
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
5.7.3Optionalhandcontrol
ForCBJCareaoptionalhandcontrolwithbatteryindicator
isavailable.
Battery
Indicator
Type
Battery
State
Description
Full
Charge
ThebatteryisOKnoneed
forcharging(100–50%).The
rightLEDisGREEN.
Partial
Charge
Thebatteryneedstobe
charged(50–25%).Themiddle
LEDisYELLOW.
Low
Charge
Thebatteryneedstobe
charged(Lessthan25%).
Acousticsignalwhenabutton
ispressed.TheleftLEDis
YELLOW.
Low
Charge
(LED
ashing)
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
5.7.4Optionalbatterycharger
(forcontrolunitswithdetachablebatteryonly)
Theprocedureofremovingorinstallingthebattery
isthesameforthecontrolunitandthecharger.
Removingthebattery
1.2.
A
B
B
1.LiftuponthehandleAonthebackofthebatteryB.
2.Liftthebatteryupandoutawayfromthecontrolunit
orcharger.
Installingthebattery
CAUTION!
Installingthebatteryimproperlymaycauseinjury
ordamage.
Makesurethereisanaudibleclickwhen
installingthebatteryonthecontrolunitor
chargertoconrmproperinstallation.
161650023-D
Usage
1.PlacethebatteryB
onthecontrolunitor
chargerasshownand
makesurethereisan
audibleclick.
B
Whenthebatteryisplacedonthecharger,thecharge
LEDwillilluminate.Whenchargingiscomplete,
chargeLEDwillswitchoff.
Abatteryneedingtobefullyrechargedwilltake
approximatelyfourhours.
1650023-D17
InvacarBirdie®EVO
6PatientTransfer
6.1Generalsafetyinformation
WARNING!
RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury
ordamage.
Beforetransferringtoastationaryobject
(wheelchair,bed,commodeorothersurface)
checkthattheweightcapacitycanwithstand
thepatientsweight.
Ifapplicablethewheelorcastorlocksofthe
stationaryobject(wheelchair,bedetc.)mustbe
inalockedpositionbeforeloweringthepatient
ontoorliftingthepatientoff.
WARNING!
RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury
ordamage.
Usethepushbarsonthemastatalltimesto
pushorpullthelift.
Avoidusingtheliftonanincline.Invacare
recommendsthattheproductonlybeusedon
aatsurface.
Duringtransfer,withthepatientsuspendedin
aslingattachedtothelift,DONOTrollliftover
unevensurfacesthatcouldcausethelifttotip
over.
WARNING!
RiskofInjuryorDamage
Damagetopartsofthelift(handcontrol,castors,
etc.)causedbyimpactwiththeoor,wallsor
otherstationaryobjectsmaycausedamageto
theproductandleadtoinjury.
DONOTallowpartsofthelifttoimpactthe
oor,wallsorotherstationaryobjects.
ALWAYSstorethehandcontrolproperlywhen
notinuse.
WARNING!
RiskofInjury
Thespreaderbarcanmovesuddenlyandcause
injury.
Whenpositioninglift,beawareoftheposition
ofthespreaderbarandthepatient.
WARNING!
RiskofEntrapmentorStrangulation
Thehandcontrolcordcancauseinjuryif
improperlypositionedandsecured.
ALWAYSbeawareofthelocationofthehand
controlcordinrelationtothepatientand
caregivers.
DONOTallowthehandcontrolcordtobecome
entangledaroundthepatientandcaregivers.
Thehandcontrolmustbesecuredproperly.
ALWAYSstorethehandcontrolproperlywhen
notinuse.
WARNING!
RiskofEntrapmentorStrangulation
Itemsinthepatientssurroundingscancause
entrapmentstrangulationduringlifting.Toavoid
entrapmentorstrangulation:
Beforelifting,checkthatthepatientis
completelyfreeofhis/hersurroundings.
WARNING!
RiskofEntrapment
Thereisariskofentrapmentbetweenthe
spreaderbarhooksandthesling.
Usecautionwhenlifting.
NEVERputhandsorngersonornearthe
hookswhenlifting.
Ensurethepatientshandsandngersareaway
fromthehooksbeforelifting.
IMPORTANT!
Alltransferproceduresdescribedinthefollowing
canbeperformedbyone(1)assistant.However,
Invacarerecommendsthattheproceduresare
performedbytwo(2)assistantswhenever
possible.
6.2Preparingtolift
WARNING!
RiskofInjury
Duringtransfersandliftoperation,theboomcan
impactthepatientorcaregiversandcauseinjury.
ALWAYSbeawareofthepositionoftheboom
duringtransfers.
Ensuretheboomispositionedinawaythatit
cannotimpactthepatientorbystanders.
ALWAYSbeawareofyourbodypositionin
relationtotheboomduringtransfers.
1.Beforeproceeding,reviewtheinformationandobserve
allwarningsindicatedin2Safety,page5andLifting
thePatient.
2.Positionthepatientontothesling.Refertoyoursling
usermanual.
3.Unlocktherearcastors.See5Usage,page13.
4.Openthelegsofthelift.See5Usage,page13.
5.Usethepushbarstomovethepatientliftintoposition.
WARNING!
Whenusingtheliftinconjunctionwith
bedsorwheelchairs,beawareofthe
positionoftheliftinrelationshiptothose
otherdevicessothattheliftdoesnot
becomeentangled.
Beforepositioningthelegsofthepatient
liftunderabed,makesurethatthearea
isclearofanyobstructions.
6.Lowerthepatientliftforeasyattachmentofthesling.
7.Locktherearcastors.See5Usage,page13
8.Attachthesling.See6.2.1Attachingtheslingtothe
lift,page19
181650023-D
PatientTransfer
6.2.1Attachingtheslingtothelift
WARNING!
RiskofInjury
Usageofwrongordamagedslingscancausethe
patienttofallorcauseinjurytoassistants.
UseanInvacareapprovedslingthatis
recommendedbytheindividual’sdoctor,nurse
ormedicalassistantforthecomfortandsafety
oftheindividualbeinglifted.
Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction
withInvacarepatientlifts.
Aftereachlaundering(inaccordancewith
instructionsonthesling),inspectsling(s)for
wear,tear,andloosestitching.
Bleached,torn,cut,frayed,ordamagedslings
areunsafeandcouldresultininjury.Discard
immediately.
DONOTalterslings.
WARNING!
RiskofInjury
Improperlyattachedoradjustedslingscancause
thepatienttofallorcauseinjurytoassistants.
Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to
ensurethatitisproperlyattachedbeforethe
patientisremovedfromastationaryobject
(bed,chairorcommode).
ThespreaderbarMUSTbeattachedtothelift
BEFOREattachingthesling.
DONOTuseanykindofplasticback
incontinencepadorseatingcushionbetween
patientandslingmaterialthatmaycausethe
patienttoslideoutoftheslingduringtransfer.
Makesurethatthereissufcientheadsupport
whenliftingapatient.
Positionthepatientintheslingasdirectedby
theinstructionsprovidedwiththesling.
Adjustmentsforpatientsafetyandcomfort
shouldbemadebeforemovingthepatient.
Thestrapsoftheslingmaybeequippedwithcolorcoded
loopsprovidingdifferentlengthstoplacethepatientinto
differentpositions.Shorterstrapsattheshoulderswill
produceamoreverticallift,whichwillassistwithpositioning
intoachairorwheelchair.Bylengtheningthestrapsatthe
shoulders,amorereclinedpositioncanbeachievedwhich
ismoresuitableforachairtobedtransfer.Alwaysmatch
theloopcolorsofthecorrespondingstrapsoneachsideof
theslingforanevenliftofthepatient.
1.
2.
A
B
B
A
1.PlacethedesiredloopofstrapAoverthenoseofhook
B.
2.PulldownthestrapAuntiltheloopisfullyseatedat
thebottomofhookB.
3.Repeatthestepsforeachoftheremainingstrapsof
thesling.
a.
CD
C
D
Ona2-pointspreaderbar,attachtheshoulderstrap
CbeforethelegstrapDoneachofthehooks.
b.
Ona4-pointspreaderbar,attacheachofthestraps
toanindividualhook.
6.3Transferringapatientfromabed
1.Preparetolift.See6.2Preparingtolift,page18.
2.Attachtheslingtothelift.See6.2.1Attachingthesling
tothelift,page19.
3.Unlocktherearcastors.
4.
Lowerthebedtothelowestposition.
1650023-D19
InvacarBirdie®EVO
5.
Liftthepatienthighenoughtoclearthestationary
objectwiththeirweightfullysupportedbythelift.
TheboomwillstayinpositionuntiltheDOWN
buttonispressed.
6.
Beforemovingthepatient,checkagaintomakesure
thattheslingisproperlyattachedtothehooksofthe
spreaderbar.See6.2.1Attachingtheslingtothelift,
page19.Ifanyattachmentsarenotproperlyinplace,
lowerthepatientbackontothestationaryobjectand
correcttheproblem.
7.
Usingthepushbars,movetheliftawayfromthe
stationaryobject.
8.
Usingthehandlesonthesling,turnthepatientsothat
he/shefacestheassistantoperatingthepatientlift.
9.
Lowerthepatientsothathis/herfeetrestonthebase
ofthelift,straddlingthemast.
Thelowercenterofgravityprovidesstability
makingthepatientfeelmoresecureandthelift
easiertomove.
10.
Movethepatientliftwithbothhandsrmlyonthe
pushbars.
11.Proceedwiththetransfertoanotherrestingsurfaceor
stationaryobject.
6.4Transferringapatienttoabed
1.Performthefollowingstepsinadditiontothoserequired
toliftthepatientofanotherrestingsurfaceorstationary
object
2.
Raiseorlowerthelifttopositionthepatientoverthe
stationarysurface.Besuretoraiseorlowerthepatient
enoughtoclearthesidesofthestationaryobject.
3.
Lowerthepatientontothestationarysurface.
4.Ensurethepatientisfullysupportedbythesurfaceto
whichyouaretransferring.
5.Locktherearcastors.
6.Raisethebedtoagoodworkingheight(usuallyhip
heightofcaregivers).
201650023-D
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238

Invacare Birdie EVO Benutzerhandbuch

Typ
Benutzerhandbuch
Dieses Handbuch eignet sich auch für